RecallWatchUSA

Home / Recalls / Medical Devices

Edwards Lifesciences, LLC recalls Oxymetry Cable, Rxonly. REF/UDI: HEMOXSC100/(01)00690103196986 - Product Usage: The HemoSphere oximetry cable is a multi

Recalled May 20, 2020 · FDA recall Z-1928-2020 · Edwards Lifesciences

Hazard
Intermittent communication due to inconsistent crimping of connectors onto cables. · FDA Class II
Units
35
Sold at
Worldwide distribution - US Nationwide distribution including in the states of AL, CA, FL, IL, IN and MO, and the countries of Germany, Netherlands, United Kingdom.

Description

Intermittent communication due to inconsistent crimping of connectors onto cables.

Product
Oxymetry Cable, Rxonly. REF/UDI: HEMOXSC100/(01)00690103196986 - Product Usage: The HemoSphere oximetry cable is a multi-use device that connects with a compatible Edwards monitor on one end and any approved Edwards oximetry catheter on the other end to continuously measure venous oxygen saturation.
Manufacturer(s)
Edwards Lifesciences, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov