RecallWatchUSA

Home / Recalls / Medical Devices

Elekta Inc recalls Elekta ERGO++ Product Usage: ERGO is often used for stereotactic treatments.

Recalled December 20, 2017 · FDA recall Z-0234-2018 · Elekta

Hazard
Incorrect DICOM mapping of the exported collimator or couch angles from ERGO, which would lead to incorrect rotation of the collimator or couch when using a MLC device for planning. · FDA Class II
Units
182
Sold at
Worldwide Distribution - US Nationwide in the states of: VA, CA, TX, MO, NY, WA, NM, MA, FL, CT, OK, SC, MS, NY, GA, NC and the countries of: AUSTRALIA, BANGLADESH, BRAZIL, BULGARIA, CANADA, COLOMBIA, CZECH REPUBLIC, ECUADOR, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ITALY, JAPAN, JORDAN, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, PERU, PHILIPPINES, PORTUGAL, RUSSIA, SAUDI ARABIA, SINGAPORE, SOUTH KOREA, SPAIN, TAIWAN, THAILAND, TURKEY, VENEZUELA, and VIETNAM

Description

Incorrect DICOM mapping of the exported collimator or couch angles from ERGO, which would lead to incorrect rotation of the collimator or couch when using a MLC device for planning.

Product
Elekta ERGO++ Product Usage: ERGO is often used for stereotactic treatments.
Manufacturer(s)
Elekta Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov