RecallWatchUSA

Home / Recalls / Medical Devices

Elekta, Inc. recalls Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is re

Recalled June 5, 2013 · FDA recall Z-1402-2013 · Elekta

Hazard
Unexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0. · FDA Class II
Units
31
Sold at
USA Nationwide including the states of CA, FL, IL, MD, MA, NJ, NY, NC, ND, OH, OR, PA, TN, TX, VA, WA, DC and Puerto Rico

Description

Unexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0.

Product
Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed.
Manufacturer(s)
Elekta, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Elekta, Inc. recalls Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is re | RecallWatchUSA