RecallWatchUSA

Home / Recalls / Medical Devices

Elekta, Inc. recalls MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

Recalled January 21, 2026 · FDA recall Z-1078-2026 · Elekta

Hazard
Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice · FDA Class II
Units
1
Sold at
US Nationwide distribution in the state of TX.

Description

Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice

Product
MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
Manufacturer(s)
Elekta, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Elekta, Inc. recalls MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO) | RecallWatchUSA