Home / Recalls / Medical Devices
Elekta, Inc. recalls RT Desktop The EOS RT Desktop, as with the predicate SL/SLi Series user interface, is intended to be used as the user
Recalled March 12, 2014 · FDA recall Z-1081-2014 · Elekta
Hazard
During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value. · FDA Class II
Units
489
Sold at
Worldwide Distribution - USA including AL, CA, CT, GA, FL, IL, IN, KS, KY, LA, MD, MA, MI, MN, MO, NJ, NY, OH, PA, TN, TX, VA, and Internationally to Argentina, Austria, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cuba, Czech Republic, Egypt, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Lebanon, Mexico, Morocco, Netherland, Nigeria, Pakistan, Panama, Peru, Philippines, Portugal, Russia, Serbia, Solvenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey, United Kingdom, Venezuela, Vietnam, and Yemen.
Description
During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value.
Product
RT Desktop The EOS RT Desktop, as with the predicate SL/SLi Series user interface, is intended to be used as the user interface for the SL/SLi Series Linear Accelerators that are used for the radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.
Manufacturer(s)
Elekta, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →