Encore Medical, Lp recalls EMPOWR Locking Femoral Impactor , Part # 800-05-035, packaged in instrument tray FA EMP PAT TOOL KIT Part # 800-99-096

Recalled December 18, 2019 · FDA recall Z-0611-2020 · Encore Medical

Hazard
The firm has received reports/complaints concerning the breakage of the tips of the locking femoral Impactor. Additionally, if the impactor is not seated correctly on the implant prior to impaction, this may cause off axis loading which can increase the risk of device failure. This could cause a delay in surgery. If the tip is not seen during surgery, long term risks may include an inflammatory response, infection, and/ or pain · FDA Class II
Units
410 devices
Sold at
CO, WA, CA, UT, CA, TN, TX, MS, MO, AR, OK, LA, MN, IL, IN, KY, OH, MI, PA, SC, FL, PR, AL, GA, NM, AZ, NV, VA, RI, MD, ME PA, NY OUS: None

Description

The firm has received reports/complaints concerning the breakage of the tips of the locking femoral Impactor. Additionally, if the impactor is not seated correctly on the implant prior to impaction, this may cause off axis loading which can increase the risk of device failure. This could cause a delay in surgery. If the tip is not seen during surgery, long term risks may include an inflammatory response, infection, and/ or pain

Product
EMPOWR Locking Femoral Impactor , Part # 800-05-035, packaged in instrument tray FA EMP PAT TOOL KIT Part # 800-99-096