RecallWatchUSA

Home / Recalls / Medical Devices

Encore Medical, Lp recalls LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up

Recalled February 22, 2017 · FDA recall Z-1191-2017 · Encore Medical

Hazard
Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization. · FDA Class II
Units
13 devices
Sold at
US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI

Description

Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization.

Product
LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a proper neck by means of a Morse taper stabilized during the implantation phase by a safety screw. This system is particularly indicated for revision surgery on both uncemented and cemented femoral implants, when there is significant bone loss and an abnormal meta-epiphyseal anatomy of the femur.
Manufacturer(s)
Encore Medical, Lp

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Encore Medical, Lp recalls LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up | RecallWatchUSA