Enztec Limited recalls REF 800-05-035, ENZTEC FEMORAL IMPACTOR - DJO, (01) 09421028110624 (10) 999999, To be used only with DJO EMPOWR 3D KNEE

Recalled February 3, 2021 · FDA recall Z-0612-2020 · Enztec

Hazard
The femoral impactors have a higher than normal risk of the arm breaking. During impaction the tip of the arm may fracture and separate from the instrument. This may cause a minor delay in surgery. If not retrieved, it may cause an inflammatory response that may require revision surgery. · FDA Class II
Units
312 Femoral Impactors
Sold at
CA

Description

The femoral impactors have a higher than normal risk of the arm breaking. During impaction the tip of the arm may fracture and separate from the instrument. This may cause a minor delay in surgery. If not retrieved, it may cause an inflammatory response that may require revision surgery.

Product
REF 800-05-035, ENZTEC FEMORAL IMPACTOR - DJO, (01) 09421028110624 (10) 999999, To be used only with DJO EMPOWR 3D KNEE
Manufacturer(s)
Enztec Limited
Made in
New Zealand

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov