Epitope Diagnostics, Inc. recalls EDI Novel Coronavirus COVID-19 IgG ELISA Kit, REF: KT-1032 Per label: This kit is intended for qualitative detection

Recalled November 18, 2020 · FDA recall Z-0426-2021 · Epitope Diagnostics

Hazard
The use of deionized water rather than distilled water during the manufacturing process, may have resulted in accelerated decreased OD signals for the Positive Control value. · FDA Class II
Units
2886
Sold at
U.S Distribution.: NH, CO, OR, CA, FL, NY, MD, NJ, IL, NC, MI, TX, PA, MA, OH, NM, VA, SC, HI, and TN.

Description

The use of deionized water rather than distilled water during the manufacturing process, may have resulted in accelerated decreased OD signals for the Positive Control value.

Product
EDI Novel Coronavirus COVID-19 IgG ELISA Kit, REF: KT-1032 Per label: This kit is intended for qualitative detection of human anti-COVID-19 IgG antibody in human serum. This kit is to be used as an aid for the detection of novel COVID-19. Patients with suspected clustering cases require diagnosis or differential diagnosis of novel coronavirus infection.
Manufacturer(s)
Epitope Diagnostics, Inc.