RecallWatchUSA

Home / Recalls / Medical Devices

ESAOTE S.P.A. recalls Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038

Recalled March 11, 2026 · FDA recall Z-1490-2026 · ESAOTE S.P.A

Hazard
A potential weakness has been identified in the probe body, so that, in a specific production batch, liquid may leak from the terminal part of the casing near the cable. · FDA Class II
Units
5 units
Sold at
US: Nationwide distribution in the states of NC and TX.

Description

A potential weakness has been identified in the probe body, so that, in a specific production batch, liquid may leak from the terminal part of the casing near the cable.

Product
Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038
Manufacturer(s)
ESAOTE S.P.A.
Made in
Italy

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov