RecallWatchUSA

Home / Recalls / Medical Devices

Estill Medical Technologies, Inc recalls Ultra Battery 1 Model: TA-UB1 NSN: 6515-01-592-5803 The Ultra Battery 1 (TA-UB1) is a rechargeable power source for

Recalled October 29, 2014 · FDA recall Z-0101-2015 · Estill Medical Technologies

Hazard
Estill has discovered that a fully discharged Ultra Battery 1 (TA-UB1) will reach 96.9% of its capacity during a 60 minute charging period. If you have a fully discharged Ultra Battery 1 (TA-UB1), in order to achieve a 100% of charge capacity you should charge it for a period of 90 minutes. In addition, the charging voltage range of 10-36V specified on the Ultra Battery 1 (TA-UB1) label and in t · FDA Class II
Units
6,886
Sold at
Worldwide Distribution - USA (nationwide) and Internationally to Serbia.

Description

Estill has discovered that a fully discharged Ultra Battery 1 (TA-UB1) will reach 96.9% of its capacity during a 60 minute charging period. If you have a fully discharged Ultra Battery 1 (TA-UB1), in order to achieve a 100% of charge capacity you should charge it for a period of 90 minutes. In addition, the charging voltage range of 10-36V specified on the Ultra Battery 1 (TA-UB1) label and in t

Product
Ultra Battery 1 Model: TA-UB1 NSN: 6515-01-592-5803 The Ultra Battery 1 (TA-UB1) is a rechargeable power source for the Thermal Angel Blood and IV Fluid Infusion Warmer. The TA-UB1 battery was designed to provide sufficient power for the Thermal Angel to warm and deliver one liter of blood or IV fluid infusion to a trauma patient.
Manufacturer(s)
Estill Medical Technologies, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Estill Medical Technologies, Inc recalls Ultra Battery 1 Model: TA-UB1 NSN: 6515-01-592-5803 The Ultra Battery 1 (TA-UB1) is a rechargeable power source for | RecallWatchUSA