RecallWatchUSA

Home / Recalls / Medical Devices

Ethicon Endo-Surgery Inc recalls Endoscopic Articulating Linear Cutter, 38ATW35. VASCULAR/THIN 35MM The ENDOPATH ETS-Flex Endoscopic Articulating Linear

Recalled October 26, 2016 · FDA recall Z-0296-2017 · Ethicon Endo-Surgery

Hazard
Quality control identified a component issue where the pinion gear in the device could fail under extreme use cases. If this condition occurs, staples will be formed past the cut line and the device can be opened and removed from the patient. The firing stroke may be interrupted and the knife may not fully return to the home position potentially exposing the healthcare professional to a sharps injury. · FDA Class II
Units
213
Sold at
Worldwide Distribution - US including AL, AR, AZ, CA, CO,CT, DC, FL,GA, IA,IL,KS,KY,LA,MA,MD, ME, MI,MN,MO, MS, MT,NC, ND, NH,NJ, NM, NV, NY, OH, RI, TN, TX, VA, VT, WA, WI and Internationally to Brazil and Canada.

Description

Quality control identified a component issue where the pinion gear in the device could fail under extreme use cases. If this condition occurs, staples will be formed past the cut line and the device can be opened and removed from the patient. The firing stroke may be interrupted and the knife may not fully return to the home position potentially exposing the healthcare professional to a sharps injury.

Product
Endoscopic Articulating Linear Cutter, 38ATW35. VASCULAR/THIN 35MM The ENDOPATH ETS-Flex Endoscopic Articulating Linear Cutter has application in general, gynecologic, urologic, and thoracic surgery for transection, resection, and/or creation of anastomoses
Manufacturer(s)
Ethicon Endo-Surgery Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Ethicon Endo-Surgery Inc recalls Endoscopic Articulating Linear Cutter, 38ATW35. VASCULAR/THIN 35MM The ENDOPATH ETS-Flex Endoscopic Articulating Linear | RecallWatchUSA