Ethicon Endo-Surgery Inc recalls HARMONICHD 1000i Shears (36cm Shaft Length)-indicated for soft tissue incisions when bleeding control and minimal therma

Recalled September 1, 2021 · FDA recall Z-2356-2021 · Ethicon Endo-Surgery

Hazard
An internal component may be cracked and become lodged behind the energy button potentially resulting in continuous activation of the device, may result in burns · FDA Class II
Units
68960 units
Sold at
Worldwide distribution - US Nationwide distribution and the countries of Austria, Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czechia, Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, United Arab Emirates.

Description

An internal component may be cracked and become lodged behind the energy button potentially resulting in continuous activation of the device, may result in burns

Product
HARMONICHD 1000i Shears (36cm Shaft Length)-indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. Product Code: HARHD36
Manufacturer(s)
Ethicon Endo-Surgery Inc