RecallWatchUSA

Home / Recalls / Medical Devices

Ethicon Endo-Surgery Inc recalls PROXIMATE¿ PPH Hemorrhoidal Circular Stapler Product Code:PPH03 PPH03: The PROXIMATE¿ PPH Hemorrhoidal Circular Stap

Recalled January 16, 2019 · FDA recall Z-0696-2019 · Ethicon Endo-Surgery

Hazard
Hemorrhoidal Circular Stapler may have been assembled without a washer, this may potentially cause bleeding or soft tissue injury during use as the device may not fully cut when fired · FDA Class II
Units
15,264 units
Sold at
Worldwide Distribution: US (Nationwide) and to countries of: ARGENTINA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, ECUADOR, EGYPT, INDIA, INDONESIA, JAPAN, KOREA, KUWAIT, LEBANON MEXICO, NICARAGUA, PAKISTAN, SAUDI ARABIA, SINGAPORE, TAIWAN URUGUAY and VIETNAM.

Description

Hemorrhoidal Circular Stapler may have been assembled without a washer, this may potentially cause bleeding or soft tissue injury during use as the device may not fully cut when fired

Product
PROXIMATE¿ PPH Hemorrhoidal Circular Stapler Product Code:PPH03 PPH03: The PROXIMATE¿ PPH Hemorrhoidal Circular Stapler and Accessories have application throughout the anal canal to perform surgical treatment of hemorrhoidal disease
Manufacturer(s)
Ethicon Endo-Surgery Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Ethicon Endo-Surgery Inc recalls PROXIMATE¿ PPH Hemorrhoidal Circular Stapler Product Code:PPH03 PPH03: The PROXIMATE¿ PPH Hemorrhoidal Circular Stap | RecallWatchUSA