Ethicon, Inc. recalls Ethicon Dermabond Advanced Topical Skin Adhesive-topical skin adhesive to hold closed easily approximated skin edges of
Recalled February 10, 2021 · FDA recall Z-0953-2021 · Ethicon
Hazard
Product distributed with a red tape as part of the sterile packaging which could impact the seal and result in a sterility breach · FDA Class II
Units
23,580 eaches
Sold at
US Nationwide
Description
Product distributed with a red tape as part of the sterile packaging which could impact the seal and result in a sterility breach
Product
Ethicon Dermabond Advanced Topical Skin Adhesive-topical skin adhesive to hold closed easily approximated skin edges of wounds from surgical incisions, including incisions Product Code: DNX12
Manufacturer(s)
Ethicon, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →