Home / Recalls / Medical Devices
Exactech, Inc. recalls Equinoxe Anatomic Replicator Plate Intended for Primary and Reverse Total Shoulder Arthroplasty
Recalled August 8, 2012 · FDA recall Z-2115-2012 · Exactech
Hazard
Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year · FDA Class II
Units
54,877 units for all products in Recall Event
Sold at
Worldwide Distribution - USA including FL, MO, MA, AL, KY, IN, NC, LA,PA, VA, OH, NY, ME, MI, TX, KS, TN, OK, NJ, WA, CO, GA, CA, MN, IA, OR, and IL and Internationaly to Austria, Luxenbourg, Sweden, Germany, Spain, France, Greece, Italy, Nederland, United Kingdom, Canada, India, Australia, Switzerland, and Sao Palo.
Description
Exactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that the device labeling identifies the affected parts as having a ten year shelf life; however, the packaging configuration that was used it qualified to a five year
Product
Equinoxe Anatomic Replicator Plate Intended for Primary and Reverse Total Shoulder Arthroplasty
Manufacturer(s)
Exactech, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →