Exactech, Inc. recalls Exactech Equinoxe CAGE GLENOID,POSTERIOR AUGMENT, CEMENTED, Right, Mates with: a) 38, 41, 44, 47 head, Small, Item Nu

Recalled April 10, 2024 · FDA recall Z-1415-2024 · Exactech

Hazard
The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH). · FDA Class II
Units
4007 units
Sold at
Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand

Description

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Product
Exactech Equinoxe CAGE GLENOID,POSTERIOR AUGMENT, CEMENTED, Right, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-13-32, b) 41, 44, 47, 50, head, Medium, Item Number 314-13-33, c) 44, 47, 50, 53 head, Large, Item Number 314-13-34, d) 44, 47, 50, 53 head, Extra Large, Item Number 314-13-35; Shoulder Arthroplasty
Manufacturer(s)
Exactech, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Exactech, Inc. recalls Exactech Equinoxe CAGE GLENOID,POSTERIOR AUGMENT, CEMENTED, Right, Mates with: a) 38, 41, 44, 47 head, Small, Item Nu | RecallWatchUSA