RecallWatchUSA

Home / Recalls / Medical Devices

Exactech, Inc. recalls EXACTECH Equinoxe Modular Reverse Pilot-Tip Reamer, Catalog #: 321-25-38, Catalog #: 321-25-42 and Catalog #: 321-25-46

Recalled July 12, 2017 · FDA recall Z-2666-2017 · Exactech

Hazard
Instructions for use (IFU) is updated to clarify the proper technique and instrumentation to ream the Glenoid, i.e., to include a caution statement that the surgeon should ream on-axis and that reaming off-axis may result in instrument breakage. · FDA Class II
Units
2,069
Sold at
AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, Hawaii and Puerto Rico Australia, Austria, Brazil, Canada, China, France, Germany, Greece, India, Italy, Korea, Luxembourg, Singapore, South Africa, Spain, Sweden, Switzerland, The Netherlands and United Kingdom

Description

Instructions for use (IFU) is updated to clarify the proper technique and instrumentation to ream the Glenoid, i.e., to include a caution statement that the surgeon should ream on-axis and that reaming off-axis may result in instrument breakage.

Product
EXACTECH Equinoxe Modular Reverse Pilot-Tip Reamer, Catalog #: 321-25-38, Catalog #: 321-25-42 and Catalog #: 321-25-46
Manufacturer(s)
Exactech, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Exactech, Inc. recalls EXACTECH Equinoxe Modular Reverse Pilot-Tip Reamer, Catalog #: 321-25-38, Catalog #: 321-25-42 and Catalog #: 321-25-46 | RecallWatchUSA