Home / Recalls / Medical Devices
Exactech, Inc. recalls Exactech Shoulder GPS Impactor Handle, Catalog #531-07-05.
Recalled July 5, 2023 · FDA recall Z-2050-2023 · Exactech
Hazard
The ball and spring housed within the locking mechanism of the handle can dislodge during disassembly or during impaction. · FDA Class II
Units
250 devices
Sold at
Distribution was made to AR, CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, ME, MN, NH, NV, NY, OH, OK, PA, SC, TN, TX, WI, as well as PR. There was no government/military distribution. Foreign distribution was made to Australia, France, and Japan.
Description
The ball and spring housed within the locking mechanism of the handle can dislodge during disassembly or during impaction.
Product
Exactech Shoulder GPS Impactor Handle, Catalog #531-07-05.
Manufacturer(s)
Exactech, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →