Home / Recalls / Medical Devices
Ferno-Washington Inc recalls Wheel caster set screw for Ferno iNX Wheeled Stretcher. Ferno iNX Wheeled Stretcher for pre-hospital patient transport.
Recalled September 2, 2015 · FDA recall Z-2461-2015 · Ferno-Washington
Hazard
The firm discovered an improperly installed set screw in one caster arm assembly on a unit in production. The screw assists in preventing the bearing cup on the caster assembly from backing out of the caster arm during use. · FDA Class II
Units
26 units
Sold at
Distributed in the states of AZ, CO, IA, MD, OH, PA & WA, and the countries of Australia, Canada & United Kingdom.
Description
The firm discovered an improperly installed set screw in one caster arm assembly on a unit in production. The screw assists in preventing the bearing cup on the caster assembly from backing out of the caster arm during use.
Product
Wheel caster set screw for Ferno iNX Wheeled Stretcher. Ferno iNX Wheeled Stretcher for pre-hospital patient transport.
Manufacturer(s)
Ferno-Washington Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →