Home / Recalls / Medical Devices
Flower Orthopedics Corporation recalls T25 Cannulated Screwdrivers (CSD 025) packaged in Guide Wire Kits (GWK 300) - Product Usage: Intended to be used to inse
Recalled June 9, 2021 · FDA recall Z-1698-2021 · Flower Orthopedics
Hazard
Cannulated Screwdrivers may break at the driver tip along the shaft, near the handle extending surgery time and additional x-ray to identify the fragments from the breakages · FDA Class II
Units
47 units
Sold at
US Nationwide distribution in the states of AZ, CO, FL, GA, IL, IN, PA, NC, VA.
Description
Cannulated Screwdrivers may break at the driver tip along the shaft, near the handle extending surgery time and additional x-ray to identify the fragments from the breakages
Product
T25 Cannulated Screwdrivers (CSD 025) packaged in Guide Wire Kits (GWK 300) - Product Usage: Intended to be used to insert screws with T25 driving features, which includes screws with diameters of 5.5mm, 6.5mm, and 7.3mm Product Code: GWK 300
Manufacturer(s)
Flower Orthopedics Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →