RecallWatchUSA

Home / Recalls / Medical Devices

Fresenius Medical Care Holdings, Inc. recalls AquaBplus and AquaBplus B2 Water Purification System. For use with a hemodialysis system to remove organic and inorgani

Recalled October 21, 2020 · FDA recall Z-0098-2021 · Fresenius Medical Care Holdings

Hazard
Incorrectly crimped connectors were installed on the main protective earth conductor on these machines. If the machine is moved or maintenance is performed, these crimp connections could become loose and may adversely affect the electrical safety of the products, potentially resulting in electrical shock to the Biomed Technician or clinic staff. · FDA Class II
Units
15 units
Sold at
US distribution to DE, NC, OK, PA, and TX.

Description

Incorrectly crimped connectors were installed on the main protective earth conductor on these machines. If the machine is moved or maintenance is performed, these crimp connections could become loose and may adversely affect the electrical safety of the products, potentially resulting in electrical shock to the Biomed Technician or clinic staff.

Product
AquaBplus and AquaBplus B2 Water Purification System. For use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from the water used for hemodialysis or related therapies. AquaBplus 1500 - Typ USA - 3 x 208V / 60Hz -Fresenius Part Number: 24-1500-1 AquaBplus B2 2000 / B1 2500 (Bplus DUO 2000) - 3 x 208V / 60Hz- Fresenius Part Number: 24-2000-0
Manufacturer(s)
Fresenius Medical Care Holdings, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Fresenius Medical Care Holdings, Inc. recalls AquaBplus and AquaBplus B2 Water Purification System. For use with a hemodialysis system to remove organic and inorgani | RecallWatchUSA