Home / Recalls / Medical Devices
Fresenius Medical Care Holdings, Inc. recalls AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Ty
Recalled July 20, 2022 · FDA recall Z-1364-2022 · Fresenius Medical Care Holdings
Hazard
AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation · FDA Class II
Units
2 units
Sold at
US Nationwide distribution in the states of NC and WV.
Description
AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation
Product
AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)
Manufacturer(s)
Fresenius Medical Care Holdings, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →