Fresenius Medical Care Holdings, Inc. recalls Blood Pressure Module BPM3 TM-2917B (Spare Part Numbers 191378 or 191389) with firmware version S005 installed in 2008 S

Recalled September 18, 2024 · FDA recall Z-3124-2024 · Fresenius Medical Care Holdings

Hazard
Machines equipped with BPM3 Blood Pressure Modules (Blood pressure monitor TM-2917B with firmware version S005) will not record Blood Pressure measurements and does not alarm if the cuff tubing is kinked close to the cuff. · FDA Class II
Units
3,718 systems
Sold at
US Nationwide. Canada.

Description

Machines equipped with BPM3 Blood Pressure Modules (Blood pressure monitor TM-2917B with firmware version S005) will not record Blood Pressure measurements and does not alarm if the cuff tubing is kinked close to the cuff.

Product
Blood Pressure Module BPM3 TM-2917B (Spare Part Numbers 191378 or 191389) with firmware version S005 installed in 2008 Series Hemodialysis Machines with Part Numbers 190573, 190713, 190766, 190858, 190895, 190908, 191124, 191126, 191128, and 191130
Manufacturer(s)
Fresenius Medical Care Holdings, Inc.