RecallWatchUSA

Home / Recalls / Medical Devices

Fresenius Medical Care Holdings, Inc. recalls Naturalyte Liquid Acid concentrate bottles, Product Code 08-2301-3 - Product Usage: used in conjunction with Fresenius R

Recalled January 20, 2021 · FDA recall Z-0835-2021 · Fresenius Medical Care Holdings

Hazard
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification. · FDA Class II
Units
1426 cases
Sold at
US Nationwide distribution.

Description

The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.

Product
Naturalyte Liquid Acid concentrate bottles, Product Code 08-2301-3 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
Manufacturer(s)
Fresenius Medical Care Holdings, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Fresenius Medical Care Holdings, Inc. recalls Naturalyte Liquid Acid concentrate bottles, Product Code 08-2301-3 - Product Usage: used in conjunction with Fresenius R | RecallWatchUSA