RecallWatchUSA

Home / Recalls / Medical Devices

Fresenius Medical Care Renal Therapies Group, LLC recalls 2008 K2 Hemodialysis Machine with software version 5.40, Models: (1) Hemodialysis SYS OLC/Diasafe PLS 190610, UDI 008408

Recalled April 11, 2018 · FDA recall Z-1278-2018 · Fresenius Medical Care Renal Therapies Group

Hazard
When the recirculation ultrafiltration (UF) Goal is set to a value greater than 200ml in service mode and the user starts a treatment using the SLED (Sustained Low Efficiency Dialysis) program the display will show an invalid message and the UF pump will run at the recirculation UF rate which may be up to 4000ml/hour. The SLED program is required to limit the UF rate to a maximum of 1000 ml/hour. · FDA Class II
Units
127
Sold at
US Distribution to the states of: AR, CA, CO, FL, GA, IL, MA, MD, MN, NC, NY, OH, PA, SC, TN, and TX.

Description

When the recirculation ultrafiltration (UF) Goal is set to a value greater than 200ml in service mode and the user starts a treatment using the SLED (Sustained Low Efficiency Dialysis) program the display will show an invalid message and the UF pump will run at the recirculation UF rate which may be up to 4000ml/hour. The SLED program is required to limit the UF rate to a maximum of 1000 ml/hour.

Product
2008 K2 Hemodialysis Machine with software version 5.40, Models: (1) Hemodialysis SYS OLC/Diasafe PLS 190610, UDI 00840861100859 (+Serial Number), (2) Machine Short Cap, OLC.DP HP 190630, UDI 00840861100866 (+Serial Number) ndicated for acute and chronic dialysis therapy.
Manufacturer(s)
Fresenius Medical Care Renal Therapies Group, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Fresenius Medical Care Renal Therapies Group, LLC recalls 2008 K2 Hemodialysis Machine with software version 5.40, Models: (1) Hemodialysis SYS OLC/Diasafe PLS 190610, UDI 008408 | RecallWatchUSA