RecallWatchUSA

Home / Recalls / Medical Devices

Fresenius Medical Care Renal Therapies Group, LLC recalls Fresenius 2008T Series Hemodialysis System

Recalled January 25, 2017 · FDA recall Z-1026-2017 · Fresenius Medical Care Renal Therapies Group

Hazard
When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine · FDA Class II
Units
65361 units (US) 249 units (OUS)
Sold at
Nationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam

Description

When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine

Product
Fresenius 2008T Series Hemodialysis System
Manufacturer(s)
Fresenius Medical Care Renal Therapies Group, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Fresenius Medical Care Renal Therapies Group, LLC recalls Fresenius 2008T Series Hemodialysis System | RecallWatchUSA