Home / Recalls / Medical Devices
Fresenius Medical Care Renal Therapies Group, LLC recalls Fresenius Modified Mixer Motor (for 132 Gallon Dissolution Units), P/N 160075. For Use in the Preparation of acid con
Recalled November 25, 2015 · FDA recall Z-0296-2016 · Fresenius Medical Care Renal Therapies Group
Hazard
Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire. · FDA Class II
Units
992 units
Sold at
Nationwide Distribution.
Description
Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.
Product
Fresenius Modified Mixer Motor (for 132 Gallon Dissolution Units), P/N 160075. For Use in the Preparation of acid concentrate for hemodialysis.
Manufacturer(s)
Fresenius Medical Care Renal Therapies Group, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →