Home / Recalls / Medical Devices
Fujifilm Medical Systems U.S.A., Inc. recalls Aspire HD (for FDR-1000AWS) The Fujifilm Digital Mammography System, Aspire HD (FDR MS-l000), generates full-field di
Recalled February 28, 2018 · FDA recall Z-0659-2018 · Fujifilm Medical Systems U.S.A
Description
FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR¿1000AWS, FDR-2000AWS, FDR-3000AWS, and CR¿IR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →