Fujifilm Medical Systems U.S.A., Inc. recalls Fujifilm FDR Go PLUS, Mobile X-Ray System Product Usage: This device is intended to be used to take a film, CR or g

Recalled March 27, 2019 · FDA recall Z-1040-2019 · Fujifilm Medical Systems U.S.A

Hazard
FUJIFILM Medical Systems U.S.A., Inc., Modality Systems became aware of the possibility that the braking mechanism might not be engaged in all cases when the driving handle has been released. · FDA Class II
Units
56 units
Sold at
US Nationwide Distribution

Description

FUJIFILM Medical Systems U.S.A., Inc., Modality Systems became aware of the possibility that the braking mechanism might not be engaged in all cases when the driving handle has been released.

Product
Fujifilm FDR Go PLUS, Mobile X-Ray System Product Usage: This device is intended to be used to take a film, CR or general radiography of patients in the hospital who cannot move and/or of outpatients in emergency. Exposure can be captured on film, CR or Digital Radiography (DR)>
Manufacturer(s)
Fujifilm Medical Systems U.S.A., Inc.