Home / Recalls / Medical Devices
Fujifilm Medical Systems U.S.A., Inc. recalls Fujinon/Fujifilm Processor EPX 2500 Operation Manual. For use in endoscopic procedures.
Recalled September 30, 2015 · FDA recall Z-2805-2015 · Fujifilm Medical Systems U.S.A
Hazard
The EPX-2500 Operation Manual and sales brochure incorrectly identify that the ED-530XT endoscope is meant to be used with the EPX-2500 Processor. The Operation Manual and sales brochure state that when the ED-530XT endoscope is used with the 2500 processor, a "super image" is displayed when it is actually a "standard image" that is displayed. · FDA Class II
Units
60
Sold at
Nationwide US and Canada
Description
The EPX-2500 Operation Manual and sales brochure incorrectly identify that the ED-530XT endoscope is meant to be used with the EPX-2500 Processor. The Operation Manual and sales brochure state that when the ED-530XT endoscope is used with the 2500 processor, a "super image" is displayed when it is actually a "standard image" that is displayed.
Product
Fujinon/Fujifilm Processor EPX 2500 Operation Manual. For use in endoscopic procedures.
Manufacturer(s)
Fujifilm Medical Systems U.S.A., Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →