Home / Recalls / Medical Devices
GE Healthcare Biosciences recalls DeltaVision OMX SR Imaging System Model Number: 29115476 The DeltaVision OMX SR Imaging System is a piece of laborat
Recalled November 28, 2018 · FDA recall Z-0388-2019 · GE Healthcare Biosciences
Hazard
Non-conformance with the DeltaVision OMX SR, the interlocks are not wired in a redundant fashion to make the interlock system single-fault tolerant. · FDA Class II
Units
39
Sold at
Worldwide - US Nationwide Distribution
Description
Non-conformance with the DeltaVision OMX SR, the interlocks are not wired in a redundant fashion to make the interlock system single-fault tolerant.
Product
DeltaVision OMX SR Imaging System Model Number: 29115476 The DeltaVision OMX SR Imaging System is a piece of laboratory equipment for super resolution imaging offluorescently-la beling biological samples. Techniques such as 3D Structured Illumination, Total Internal Reflection Fluorescence and localization microscopy may be used to examine a variety of samples over long periods of time.
Manufacturer(s)
GE Healthcare Biosciences
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →