GE Healthcare Finland Oy recalls GE Healthcare CARESCAPE Monitor B850 Product Usage: Intended uses of CARESCAPE B850 V2 in 510(k) countries [UR_CRS

Recalled October 17, 2018 · FDA recall Z-0096-2019 · GE Healthcare Finland Oy

Hazard
When multiple CARESCAPEMonitor B450 or B850 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected. · FDA Class II
Units
47,937 units
Sold at
Worldwide Distribution-US Nationwide

Description

When multiple CARESCAPEMonitor B450 or B850 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected.

Product
GE Healthcare CARESCAPE Monitor B850 Product Usage: Intended uses of CARESCAPE B850 V2 in 510(k) countries [UR_CRSCP-1046]: The CARESCAPE Monitor B850 is a multi-parameter high acuity patient monitor intended for use in multiple areas within a professional healthcare facility. The CARESCAPE Monitor B850 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B850 is indicated for monitoring of: -hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation). -Respiratory (impedance respiration, airway gases (CO2, O2, N2O and anesthetic agents), spirometry, gas exchange) and -neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS) and neuromuscular transmission. The CARESCAPE Monitor B850 also provides alarms, trends, snapshots and events, and calculations and can be connected to displays, printers and recording devices.
Manufacturer(s)
GE Healthcare Finland Oy
Made in
Finland