GE Healthcare recalls GE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR220 amx
Recalled May 6, 2015 · FDA recall Z-1407-2015 · GE Healthcare
Hazard
An issue was identified with the mAs accuracy check performed on the system that could result in inaccuracies in the 10% to 20% range for exposure times longer than 15 msec. · FDA Class II
Units
6 units in US
Sold at
US Distribution to the states of: AZ, FL, GA, MO and WI.
Description
An issue was identified with the mAs accuracy check performed on the system that could result in inaccuracies in the 10% to 20% range for exposure times longer than 15 msec.
Product
GE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR220 amx
Manufacturer(s)
GE Healthcare