Home / Recalls / Medical Devices
GE Healthcare recalls GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model/Catalog/Code Numbers: a)¿¿¿¿¿ 2089507-120; b
Recalled October 16, 2024 · FDA recall Z-0039-2025 · GE Healthcare
Hazard
GE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) version v6.0 SP11.x where the latest report is not shown by default to the user. The issue may occur when a study has multiple DICOM structured reports (i.e. preliminary, final, and/or addendum) stored in different series. · FDA Class II
Units
1500 units
Sold at
Worldwide distribution.
Description
GE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) version v6.0 SP11.x where the latest report is not shown by default to the user. The issue may occur when a study has multiple DICOM structured reports (i.e. preliminary, final, and/or addendum) stored in different series.
Product
GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model/Catalog/Code Numbers: a)¿¿¿¿¿ 2089507-120; b)¿¿¿¿ 2089507-129; c)¿¿¿¿¿ 2089507-131; d)¿¿¿¿ 2089507-135; e)¿¿¿¿ 2089507-136; f)¿¿¿¿¿¿ 2089507-140; g)¿¿¿¿¿ 2089507-144; h)¿¿¿¿ 2089507-145; i)¿¿¿¿¿¿ 2089507-149; j)¿¿¿¿¿¿ 2089507-153; k)¿¿¿¿¿ 2089507-154; l)¿¿¿¿¿¿ 2089507-155; m)¿¿¿ 2089507-156; n)¿¿¿¿ 2089507-157; o)¿¿¿¿ K2037DAED; p)¿¿¿¿ K2037DBED; q)¿¿¿¿ K2037DED; r)¿¿¿¿¿ K2037DJED; s)¿¿¿¿¿ K2037DNED; t)¿¿¿¿¿ K2042TTED; u)¿¿¿¿ K2049DC; v)¿¿¿¿¿ K2049DD; w)¿¿¿ K2049DF; x)¿¿¿¿¿ P2037DAED; y)¿¿¿¿¿ P2037DBED; z)¿¿¿¿¿ P2037DED; aa)¿¿ To be provided
Manufacturer(s)
GE Healthcare
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →