RecallWatchUSA

Home / Recalls / Medical Devices

GE Healthcare, LLC recalls Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition

Recalled December 1, 2021 · FDA recall Z-0242-2022 · GE Healthcare

Hazard
The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100 times) than the actual dose delivered. · FDA Class II
Units
105 in total
Sold at
Worldwide and US Nationwide Distribution

Description

The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100 times) than the actual dose delivered.

Product
Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
Manufacturer(s)
GE Healthcare, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

GE Healthcare, LLC recalls Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition | RecallWatchUSA