Home / Recalls / Medical Devices
GE Healthcare, LLC recalls CT Goldseal Optima CT 600
Recalled August 8, 2018 · FDA recall Z-2583-2018 · GE Healthcare
Hazard
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply). · FDA Class II
Units
42 units total
Sold at
Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
Description
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Product
CT Goldseal Optima CT 600
Manufacturer(s)
GE Healthcare, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →