GE Healthcare, LLC recalls GE Healthcare Definium 5000/5220493-2 X-Ray System

Recalled March 6, 2013 · FDA recall Z-0704-2013 · GE Healthcare

Hazard
When quickly switching the protocols from AEC to Fixed under the same view, if the user selection process time is less than 0.5 seconds, the User Interface (UI) may display different exposure techniques than selected, and may not display AEC setting. The exposure can still be made as commanded, however it may not be properly indicated. · FDA Class II
Units
312 installed in US
Sold at
Nationwide Distribution

Description

When quickly switching the protocols from AEC to Fixed under the same view, if the user selection process time is less than 0.5 seconds, the User Interface (UI) may display different exposure techniques than selected, and may not display AEC setting. The exposure can still be made as commanded, however it may not be properly indicated.

Product
GE Healthcare Definium 5000/5220493-2 X-Ray System
Manufacturer(s)
GE Healthcare, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov