Home / Recalls / Medical Devices
GE Healthcare, LLC recalls GE Healthcare - Giraffe and Panda T-Piece Resuscitation System The T-piece Resuscitation System [ventilator, emergenc
Recalled February 20, 2013 · FDA recall Z-0748-2013 · GE Healthcare
Hazard
Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly. · FDA Class I
Units
276 units total, 109 of those nationwide.
Sold at
Worldwide Distribution - Worldwide Distribution - USA including AL, AZ, CA, CO, FL,GA, IA, ID, IL, KY, MA, MD, MI, MN, MO, MS, MT, NC, ND ,NE, NM, NV, NY, OH, OR, PA ,TX, VT, WA, WI and WV; and the countries of Australia, Belgium, Canada, China, France, New Zealand, Poland, Saudi Arabia, Switzerland and United Kingdom.
Description
Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.
Product
GE Healthcare - Giraffe and Panda T-Piece Resuscitation System The T-piece Resuscitation System [ventilator, emergency powered resuscitator]/Panda Bag and Mask Resuscitation System provides the basic equipment required for pulmonary resuscitation of infants.
Manufacturer(s)
GE Healthcare, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →