GE Healthcare, LLC recalls GE Healthcare Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ, Innova 2121 IQ, Innova 3131 IQ , Innova 3100, Innova 4100.

Recalled May 22, 2013 · FDA recall Z-1318-2013 · GE Healthcare

Hazard
While performing a fluoroscopic examination on the GE Innova systems, there is a potential of nonrecoverable loss of imaging modes with no advance warning, due to a relay failure in the power distribution box (PDB). Although no injuries have been reported due to this issue to date, there is risk of complications during the procedure as a result of imaging functionality loss. · FDA Class II
Units
713
Sold at
Worldwide Distribution -- Nationwide including the states of AL, AZ, AR, CA, CO, CT, DC, FL, GA, GUAM, HI, ID, IL , IN, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NH, NM, NY, NC, ND, OH, OK, OR, PA, PR, SC, TN, TX, UT, VT, VA, WA, WY and WI., and the countries of Philippines, Peru, Paraguay, Panama, Mexico, Malaysia, Korea, Japan, Guatemala, Ecuador, Dominican Republic , Columbia, China, Chile, Canada, Brazil.,Argentina and Venezuela

Description

While performing a fluoroscopic examination on the GE Innova systems, there is a potential of nonrecoverable loss of imaging modes with no advance warning, due to a relay failure in the power distribution box (PDB). Although no injuries have been reported due to this issue to date, there is risk of complications during the procedure as a result of imaging functionality loss.

Product
GE Healthcare Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ, Innova 2121 IQ, Innova 3131 IQ , Innova 3100, Innova 4100. For use in generating fluoroscopic images of human anatomy for diagnostic and interventional procedures.
Manufacturer(s)
GE Healthcare, LLC