GE Healthcare, LLC recalls GE Healthcare, Innova 2100IQ, Innova 3100IQ, Innova 4100IQ, Cardiovascular and Interventional Imaging System.
Recalled April 24, 2013 · FDA recall Z-1142-2013 · GE Healthcare
Hazard
As part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In the situation a staff member (doctor, nurse, operator, anesthetist, etc&) stands in the region between the C-arm gantry pivot and the head section of the table, there is a risk of co · FDA Class II
Units
48
Sold at
Worldwide Distribution - Nationwide Distribution including the states of FL, IL, IN, LA, MI, MN, NJ, NM, NY, OH, PR, SC, TX and VA and the countries of Vietnam, Latvia, Italy, Israel, Indonesia, Hong Kong, United Kingdom, Finland, Spain, Czech Republic, Canada, Bahrain and Australia.
Description
As part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In the situation a staff member (doctor, nurse, operator, anesthetist, etc&) stands in the region between the C-arm gantry pivot and the head section of the table, there is a risk of co
Product
GE Healthcare, Innova 2100IQ, Innova 3100IQ, Innova 4100IQ, Cardiovascular and Interventional Imaging System.
Manufacturer(s)
GE Healthcare, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →