GE Healthcare, LLC recalls GE NM/CT 850 system Model Number H3907AD - Product Usage: intended for general Nuclear Medicine imaging procedures for d

Recalled January 20, 2021 · FDA recall Z-0853-2021 · GE Healthcare

Hazard
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector. · FDA Class II
Units
24 units
Sold at
Worldwide

Description

GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.

Product
GE NM/CT 850 system Model Number H3907AD - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
Manufacturer(s)
GE Healthcare, LLC