RecallWatchUSA

Home / Recalls / Medical Devices

GE Healthcare, LLC recalls Optima NM/CT 640, Model number 3101RL - Product Usage: intended for General Nuclear Medicine imaging procedures for dete

Recalled January 20, 2021 · FDA recall Z-0858-2021 · GE Healthcare

Hazard
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector. · FDA Class II
Units
7 units
Sold at
Worldwide

Description

GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.

Product
Optima NM/CT 640, Model number 3101RL - Product Usage: intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
Manufacturer(s)
GE Healthcare, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov