GE Healthcare, LLC recalls Optima NM/CT 640 Product Usage: The GE Optima NM/CT 640 system is intended for General Nuclear Medicine imaging proc
Recalled May 8, 2019 · FDA recall Z-1266-2019 · GE Healthcare
Hazard
During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy. · FDA Class II
Units
35 units
Sold at
US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK
Description
During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
Product
Optima NM/CT 640 Product Usage: The GE Optima NM/CT 640 system is intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine (NM) system using a variety of scanning modes supported by various acquisition types, and a CT component which is intended specifically for enabling attenuation correction and anatomical localization on SPECT studies.
Manufacturer(s)
GE Healthcare, LLC