GE Healthcare, LLC recalls SIGNA Premier Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner desi
Recalled February 5, 2020 · FDA recall Z-0874-2020 · GE Healthcare
Hazard
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images. · FDA Class II
Units
62 (30 US, 32 OUS) total devices
Sold at
Worldwide distribution - US nationwide distribution and countries of Australia, China, France, Germany, India, Italy, Japan, Netherlands, Oman, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
Description
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.
Product
SIGNA Premier Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.