GE Healthcare, LLC recalls SIGNA Vibrant Nuclear Magnetic Resonance Imaging System, 1.5T Signa HDx, 1.5T Signa HDxt, Signa HDi - Product Usage: is

Recalled February 5, 2020 · FDA recall Z-0877-2020 · GE Healthcare

Hazard
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system s date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images. · FDA Class II
Units
62 (30 US, 32 OUS) total devices
Sold at
Worldwide distribution - US nationwide distribution and countries of Australia, China, France, Germany, India, Italy, Japan, Netherlands, Oman, Spain, Sweden, Switzerland, Turkey, and United Kingdom.

Description

It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system s date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.

Product
SIGNA Vibrant Nuclear Magnetic Resonance Imaging System, 1.5T Signa HDx, 1.5T Signa HDxt, Signa HDi - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
Manufacturer(s)
GE Healthcare, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov