Home / Recalls / Medical Devices
Ge Inspection Technologies Llc recalls v|tome|x L Series non-destructive testing cabinet x-ray systems
Recalled November 25, 2020 · FDA recall Z-0442-2021 · Ge Inspection Technologies
Hazard
It was discovered during testing of the v/tome/x L450 system s/n PA1790, that the pre-warning signals does not provide "Audible and visible warning signals within the cabinet" if the "warm up" mode was not completed after initializing the system when first turned on. · FDA Class II
Units
N/A
Sold at
USA Canada
Description
It was discovered during testing of the v/tome/x L450 system s/n PA1790, that the pre-warning signals does not provide "Audible and visible warning signals within the cabinet" if the "warm up" mode was not completed after initializing the system when first turned on.
Product
v|tome|x L Series non-destructive testing cabinet x-ray systems
Manufacturer(s)
Ge Inspection Technologies Llc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →