GE Medical Systems Information Technologies Inc recalls PORTRAIT CORE SERVICES SOFTWARE V1.1, Ref# 5700209

Recalled September 2, 2026 · FDA recall Z-3036-2026 · GE Medical Systems Information Technologies

Hazard
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area. · FDA Class II
Units
23 units
Sold at
US: MI, NY and Australia, Czech Republic, Finland, Germany, Italy, Netherlands, Qatar, Saudi Arabia, Switzerland, United Arab Emirates, United Kingdom.

Description

Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.

Product
PORTRAIT CORE SERVICES SOFTWARE V1.1, Ref# 5700209
Manufacturer(s)
GE Medical Systems Information Technologies Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov