RecallWatchUSA

Home / Recalls / Medical Devices

GE Medical Systems, LLC recalls B125M Patient Monitor, REF 6160000-004-XXXXXX

Recalled December 28, 2022 · FDA recall Z-0732-2023 · GE Medical Systems

Hazard
GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be inaccurate when used in a location that is not at or near sea level. · FDA Class II
Units
5962
Sold at
Worldwide distribution.

Description

GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be inaccurate when used in a location that is not at or near sea level.

Product
B125M Patient Monitor, REF 6160000-004-XXXXXX
Manufacturer(s)
GE Medical Systems, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

GE Medical Systems, LLC recalls B125M Patient Monitor, REF 6160000-004-XXXXXX | RecallWatchUSA