GE Medical Systems, LLC recalls GE Healthcare, Discovery 3.0T MR750, 5148810-2, 5373011-2, 5499460-2.
Recalled October 31, 2018 · FDA recall Z-0201-2019 · GE Medical Systems
Hazard
GE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used with the 3T MR750W Surgical Suite Scanners. Coil overheating can occur when the device is used in Mode 2 setup. This could lead to a serious patient thermal injury. There have been no injuries reported as a result of this issue. UPDATE 5/27/2016: Recall has been updated to include an additional product which includes the Discovery 3.0T MR750 systems running Application Software DV24 R01, DV25 R02, or DV25.1 M3. · FDA Class II
Units
356 (98 US; 258 OUS)
Sold at
Worldwide Distribution- US Distribution including the states of AZ, CA, CO, IL IN, MI, MN, NY, OH, OK, SD, TX, WI, VA, DE, FL, MD, ME, MO, NC, NH, OR, PA, TN, WI, DC, PR and the countries of CANADA, CHINA, CZECH REPUBLIC, ITALY, JAPAN, NETHERLANDS, NEW ZEALAND, PORTUGAL, ROMANIA, SOUTH AFRICA, SWITZERLAND, TURKEY, SWEDEN, AUSTRALIA, BRAZIL, CZECH REPUBLIC, EGYPT, FRANCE, GERMANY, ITALY, MEXICO, RUSSIA, SAUDI ARABIA, TAIWAN, UNITED KINGDOM, INDONESIA, ESTONIA, REPUBLIC OF KOREA, POLAND, KUWAIT, SPAIN.
Description
GE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used with the 3T MR750W Surgical Suite Scanners. Coil overheating can occur when the device is used in Mode 2 setup. This could lead to a serious patient thermal injury. There have been no injuries reported as a result of this issue. UPDATE 5/27/2016: Recall has been updated to include an additional product which includes the Discovery 3.0T MR750 systems running Application Software DV24 R01, DV25 R02, or DV25.1 M3.
Product
GE Healthcare, Discovery 3.0T MR750, 5148810-2, 5373011-2, 5499460-2.
Manufacturer(s)
GE Medical Systems, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →